Comparative investigation of primary stability of four different dental implants in low-density bone model


1 Department of Prosthodontics, Dental Research Center, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran
2 Department of Periodontics, Dental Implants Research Center, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran
3 Department of Prosthodontics, Faculty of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran
4 Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran

Correspondence Address:
Dr. Faezeh Atri
Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences, North Karegar Street, Tehran
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1735-3327.249557

Rights and Permissions

Background: Primary stability is one of the prerequisites of immediate loading. The aim of this study was to compare the primary stability of four different implant systems in low-density bone models. Materials and Methods: In this in vitro experimental study, 20 fixtures from four implant body designs were selected: Zimmer Tapered Screw-Vent (ZTSV), NobleReplace Tapered (NRT), Replace Select Tapered (RST), and Dentium SuperLine (DSL). Fixtures were inserted in low-density bone models according to manufacturer drilling protocol by one surgeon. Measurement of insertion torque value (ITV), resonance frequency analysis (implant stability quotient [ISQ]), and reverse torque value (RTV) was recorded for each fixture. The data were analyzed with one-way ANOVA and post hoc tests (Tukey honestly significant difference) (P < 0.05). Results: ZTSV had significantly lower amount of insertion torque in comparison to other systems (P = 0.045). RTV was significantly lower in ZTSV in comparison to DSL and NRT (P = 0.004). ISQ value in NRT (ISQ = 67) was significantly higher than other systems (P = 0.000). The lowest amount of ISQ was in RST system (ISQ = 53) with significant difference (P = 0.000). Conclusion: This study demonstrated that primary stability of different implant systems was not comparable and implant design was effective on ITV, RTV, and ISQ.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed407    
    Printed36    
    Emailed0    
    PDF Downloaded68    
    Comments [Add]    

Recommend this journal