Finite element analysis of the effect of framework materials at the bone–implant interface in the all-on-four implant system
Kasturi Chandrashekhar Kelkar, Vinaya Bhat, Chethan Hegde
Department of Prosthodontics, A B Shetty Memorial Institute of Dental Sciences, Nitte (Deemed to be University), Mangalore, Karnataka, India
Date of Submission | 05-Jan-2020 |
Date of Acceptance | 18-Mar-2020 |
Date of Web Publication | 23-Feb-2021 |
Correspondence Address: Dr. Vinaya Bhat Department of Prosthodontics, A B Shetty Memorial Institute of Dental Sciences, Deralakatte, Mangalore - 575 018, Karnataka India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/1735-3327.310031
Background: The “All-on-four” concept for treatment of edentulous arches incorporates four implants that are placed in between mental foramina in the mandible. The prosthetic framework is an important parameter in stress/strain concentration at the implants, prosthesis, and the underlying bone. Materials such as titanium, zirconia, and carbon fibers have been used for fabrication of framework in the past. The aim of this study was to analyze the effect of framework materials in the “All-on-four” implant system. Materials and Methods: Finite element three-dimensional (3D) model of edentulous mandible was simulated using a computerized tomographic scan data of an edentulous patient. Threaded implants were replicated along with the abutments using 3D modeling software and the framework was designed and simulated using material properties of titanium, zirconia, and polyetheretherketone (PEEK). Axial and nonaxial load of 200 N was applied at the abutment region of right distal implants. The computer-generated numerical values were tabulated and analysed by ANSYS software. Results: Principal strain, von Mises stress and micromotion were assessed in the peri-implant bone region to evaluate its stress condition. Zirconia framework showed the least stress/strain values at axial and oblique loading. Maximum strain values were seen at the PEEK framework material. Zirconia framework in all models showed the least micromotion/displacement. Conclusion: The stress distribution pattern at implant–bone interface was influenced by the framework material used. The framework material, loading site, and direction of forces influenced the stresses and displacement at the bone–implant interface.
Keywords: Polyetheretherketone, titanium, zirconia
How to cite this article: Kelkar KC, Bhat V, Hegde C. Finite element analysis of the effect of framework materials at the bone–implant interface in the all-on-four implant system. Dent Res J 2021;18:1 |
How to cite this URL: Kelkar KC, Bhat V, Hegde C. Finite element analysis of the effect of framework materials at the bone–implant interface in the all-on-four implant system. Dent Res J [serial online] 2021 [cited 2023 Mar 27];18:1. Available from: https://www.drjjournal.net/text.asp?2021/18/1/1/310031 |
Introduction | |  |
In edentulous patients, implant-supported full-arch restoration is a common treatment option. An implant system called “All-on-four” used for treating such patients incorporates four implants that are placed between the mental foramina region of the mandible. In this system, two implants are placed axially in the anterior region, while two are placed in the posterior region at an angle. All the four implants are connected through a superstructure framework.[1]
Materials used for the prosthetic superstructure framework play a significant role in its biomechanical success. A rigid framework allows absorption and balanced distribution of stresses to prevent deformations.[2] Materials such as titanium, zirconia, and carbon fibers have been used for the fabrication of framework in the past. Metal frameworks used for the prosthesis design present good mechanical properties. Titanium has been a practical material for the fabrication of prosthesis superstructure on implants.[1] An esthetic alternative is a zirconia framework as it is biocompatible and has good mechanical properties and high flexural strength. However, conflicting results are seen in studies evaluating load transferred by this material.[3]
Recently, a new material called polyetheretherketone (PEEK) has been introduced which exhibits excellent mechanical properties and is used in the fabrication of removable dentures, fixed restorations, dental implants, and implant abutments.[4] The Young's modulus of this material is close to that of human bone making it as elastic as a bone.[5]
Studies have suggested that stiffer materials show higher stress values in the prosthetic framework as compared to less rigid materials.[2] Materials having higher elastic modulus will resist deformation, thus increasing the stress concentration. A framework material with a lower modulus of elasticity could decrease the occlusal forces and evenly distribute load.[3] However, it has also been noted that stiffer materials transfer lower stresses to other components of the system. These biomechanical complications can compromise the osseointegration of implants to induce bone resorption.[6] Furthermore, prior studies propose that the material properties, number of implants, its distribution as well as distal support aid in determining the stress levels and displacement in an implant-supported system.[2],[3],[7]
The three-dimensional (3D) finite element method (FEM) is a numerical procedure analyzing structures using computer models of a material. It uses a complex system of points (nodes) and elements which makes a grid called a mesh. The mesh is programmed to contain the material and structural properties (Young's modulus and Poisson's ratio), and the design is stressed and analyzed for specific results.[8]
The purpose of the present study was to analyze the effects of different materials used for framework fabrication on “All-on-four” implant system using FEM. The three framework materials used in the study varied in their material properties and were critical for the evaluation of stresses in peri-implant region. Evaluation of various stress distribution patterns for axial and nonaxial stresses and micromotion that occurs around these implants was also assessed.
Materials and Methods | |  |
Designing a finite element model of “All-on-four implant” system
A 3D model of edentulous mandible was simulated using a computerized tomographic scan data of an edentulous patient. Threaded implants (11.5 mm, 13 mm length, and 4 mm diameter) were replicated along with the abutments using 3D modeling software. The bone segments were modeled as cortical bone representing the outer shell of 2-mm thickness and the inner volume of cancellous bone. The implants had 100% bone–implant contact. Following the “All-on-four” concept, two anterior implants were placed at the lower lateral incisor region, while the posterior implants were angled distally (30°) and placed anterior to the mental foramen [Figure 1].[1] | Figure 1: Mandible model with four implants according to “All-on-Four” concept.
Click here to view |
Designing of a framework using material properties of titanium, zirconia, and polyetheretherketone
The framework was designed as a solid bar of height 5 mm and width 6 mm following the shape of the mandible. The cantilever extension was 11.5 mm. At the end of the bar, 2-mm diameter circle was created on the outer occlusal surface to standardize the load application. The framework was simulated using material properties of titanium, zirconia, and PEEK, obtained from previous studies. Young's modulus and Poisson's ratio were assigned to each of the components [Figure 2]. The materials were assumed to have isotropic linear elasticity and inhomogeneity for the bone. The properties of the materials used are shown in [Table 1]. | Figure 2: Standardized load application on the outer occlusal surface of framework.
Click here to view |
Solid elements were used to represent the geometry. All the 3D models of the implants and framework were built using Solid Edge software (Siemens PLM Software, USA). Bone was built by reverse engineering processor. Meshwork was prepared using HyperMesh software. The analysis was done by ANSYS software (USA).
Application of load on the three frameworks
The axial load of 200 N was applied at the abutment region of right distal implants. Nonaxial load of 200 N was applied at an angle of 30° at the same site [Figure 3].
Evaluation of the amount of stress distribution
The software gave quantitative values at different locations. Principal compressive and tensile strain along with von Mises stress was assessed in the peri-implant bone to evaluate the stress condition of the bone.
Evaluation of micromotion
Micromotion was computed as the relative displacement between two nodes (a node of bone side and implant side) of elements on the interface.
Results | |  |
Strain distribution
The compressive and tensile strains were observed at the posterior implant and cantilever extension on axial and oblique loading. Maximum strain values were observed with the PEEK framework material. On the axial loading of the posterior implant, the tensile strain of Peek framework was 0.004888 and the compressive strain was 0.008001 [Figure 4]; [Table 2]. At the posterior implant on oblique loading, the tensile and compressive strains of the Peek framework were 0.001781 and 0.002692, respectively [Table 3]. Least tensile strain values in cantilever loading were noted at zirconia framework of 0.002656. In oblique posterior implant loading, cantilever oblique loading, and cantilever axial loading, zirconia and titanium framework showed similar strain values, indicating that the framework material and the length of cantilever influence the strain values [Table 4] and [Table 5]. | Figure 4: Compressive strain at polyetheretherketone framework on axial loading of posterior implant.
Click here to view |
 | Table 2: Stress concentration and displacement in the three framework materials on axial loading of distal implant
Click here to view |
 | Table 3: Stress concentration and displacement in the three framework materials on oblique loading of distal implant
Click here to view |
 | Table 4: Stress concentration and displacement in the three framework materials on cantilever axial loading
Click here to view |
 | Table 5: Stress concentration and displacement in the three framework materials on cantilever oblique loading
Click here to view |
von Mises stress distribution
von Mises stresses were higher on oblique loading of the titanium and PEEK framework materials in comparison to axial loading at cantilever length. The least amount of von Mises stresses was noted with the zirconia framework on axial and oblique loading of 21.0644 Mpa and 20.137Mpa, respectively [Figure 5]. PEEK framework had the highest von Mises stress values of 107.589 Mpa on axial loading and 97.4957 Mpa on oblique loading. Thus, the framework material and the site of load application influenced the von Mises stresses. Among the three materials, zirconia framework showed better stress distribution followed by a titanium framework. | Figure 5: von Mises stress at zirconia framework on oblique loading of posterior implant.
Click here to view |
Micromotion/displacement
The maximum micromotion was observed in the PEEK framework on axial loading at the posterior implant [Figure 6]. Zirconia framework in all the models showed the least micromotion compared to titanium and PEEK frameworks. The amount of micromotion seen was influenced by the loading site and the loading direction. | Figure 6: Micromotion at polyetheretherketone framework on axial loading of posterior implant.
Click here to view |
Discussion | |  |
In the present study, stress distribution and micromotion at bone–implant interface using different framework materials in “All-on-four” concept in the edentulous mandible were analyzed.
Studies have shown the occlusal masticatory forces in the posterior region of the tooth to be around 220N.[1] In the present study, a force of 200N was applied in an axial and oblique direction at the abutment region of the right distal implant and cantilever extension. The value was standardized in accordance with the previous studies.[2],[9]
The “All-on-four concept” eliminates the need for augmentation procedure in many cases. The angulation of distal implants also opens a window for placement of longer implants, thus enhancing the load distribution.[9] In the present study, a distal implant was angulated at 30°. This was in accordance with the previous finite element studies.[10],[11],[12] The distal implant angulation is said to have a dominant effect in reducing stresses at the cortical and cancellous bone. A study conducted to analyze the effect of different implant inclinations in the maxillary arch concluded that stresses in the crestal bone declined with the increase in distal implant angulation.[11] These results were not in agreement with the study conducted by Malhotra et al. who compared the stresses developed at 30° and 40° and found no statistically significant difference.[12]
The knowledge of stress distribution around the implant–bone interface is crucial for its long-term stability. The stresses transferred onto the bone from different framework materials are divergent due to the variation in its Young's modulus of elasticity.[3] Since the bone has both ductile and brittle response, the use of primary stress/strain is appropriate for evaluating the yielding failure behavior.[1] Stiffer implant materials tend to absorb the stresses being transmitted to specific bodily area. This phenomenon is referred as stress shielding where the bodily tissues are shielded resulting in adjacent bone resorption.[13] Von mises stress is a combination of normal and shear stresses which predicts the yielding of materials under complex loading.[14] Micromotion is the minute displacement at the implant–bone interface affecting the adjacent tissues.[15]
In the present study, principal tensile and compressive strains were observed at the crestal cortical bone. The length of the cantilever influenced the strain values. A study conducted by Horita et al. compared cantilever and non-cantilever loading in immediate implant placement. They found 45.3%–52.5% reduction in peak compressive strains on non-cantilever loading.[1] The axial and oblique loading of the framework in the present study derived higher strain values as opposed to loading at the posterior implant abutment region. The least tensile strain was observed in the zirconia framework on oblique loading at the posterior implant. PEEK framework material exhibited maximum compressive strain values in all situations. This could be attributed to the higher deformity of PEEK framework due to its low modulus of elasticity.
In the present study, at distal implant axial loading, the highest von Mises stress of 107.589 Mpa was noted with PEEK framework. However, the stress values were lower on the distal implant oblique loading of the three framework materials. The least stress of 20.137 Mpa was noted at zirconia framework. Thus, the framework material exhibits the ability to absorb stresses and distribute the minimal load to the adjacent bone. Overloading of the cortical bone occurs beyond 170Mpa. The value observed here is below that considered pathologic to the bone. Resorption due to excessive loading of adjacent bone could be avoided by the use of zirconia framework. The use of rigid framework material clinically would thus prevent failure of the implant support system.
The increasing anteroposterior spread of implants shortens the length of cantilever extension. The cantilever length affects the stress distribution of the underlying bone and the bone–implant interface.[16] A study reported a cantilever length of >15 mm had greater von Mises stress in buccal and lingual cortical plates.[17] The present study had a cantilever extension of 11.5 mm. Studies indicate that longer cantilever prosthesis shows higher stress concentration at the distal implant.[12],[16] Similarly, in this study, higher stress values were noted at the cantilever loading of framework. On oblique loading of the cantilever, von Mises stress was maximum for PEEK framework followed by titanium framework. Titanium material showed similar values at both axial and oblique cantilever loading. Zirconia framework showed the least stress of 51.8369 Mpa among all parameters on cantilever loading. PEEK framework showed maximum compressive strain values at axial and oblique cantilever loading. Higher compressive strain values could be due to the material properties of PEEK, thus transmitting more stresses to other system components. The least tensile strain was seen in the oblique loading of the zirconia framework. Thus, in the present study, zirconia framework showed minimal stresses at the bone–implant interface. The values are below the pathologic loading of the underlying bone.
Excessive micromotion at the implant–bone interface can have deleterious effects on the implant system. In this study, during axial loading, the least amount of displacement was observed at the zirconia framework. The titanium and PEEK materials showed similar values on axial loading. On oblique loading of the distal implant site, all the framework materials showed similar values, and the displacement was less than axial loading. In cantilever loading, maximum micromotion was observed on axial load than oblique load. Thus, according to the present study, it can be inferred that the overall loading angle and the type of loading affected the micromotion of the implant. The values did not differ significantly for cantilever and noncantilever loading. These results however differ from the study where the smallest micromotion was seen at noncantilever loading (<1/3rd) compared to cantilever loading.[18]
Prior studies suggest that lower Young's modulus of a framework material results in greater bending forces at distal implants and larger bending of the prosthesis under functional loading.[10],[19] A study evaluated the effect of three framework materials, i.e. titanium, zirconia, and PEKK for implant-supported prosthesis in the maxilla. It was observed that the stress concentration on the framework increased with an increase in elastic modulus; however, the stress transmitted to the crown was reduced.[20] Study done by Sertgöz et al. summarized that materials with lower elastic modulus did not show substantial difference in stress patterns at bone surrounding the implants.[21] In the present study, zirconia frameworks showed low-stress values in all parameters on the distal implant and cantilever loading. This may be attributed to the high Young's modulus of the material. Thus, forces acting on the framework are absorbed by the material reducing the stresses at the distal implant and surrounding bone.
One of the limitations in this study includes the materials being assumed to have isotropic linear elasticity and inhomogeneity for bones. Although not seen in a clinical scenario, these are inherent in finite element studies due to limitations in biologic simulation. The study was also conducted under unilateral static loading. The framework materials may behave differently under cyclic loading such as those occurring during chewing movements. Further clinical studies are necessary for relevance and acceptance of the findings seen in the present study.
Conclusion | |  |
Within the limitations of this study, it was concluded that:
- The stress distribution pattern at implant–bone interface was influenced by the framework material used. Zirconia framework showed minimal stress distribution in all parameters
- The site of load application and the direction of forces in axial and oblique patterns influence the compressive and tensile strain. PEEK framework material had the highest strain values
- The length of cantilever increased the stress concentration in all three framework materials
- The framework material, loading site, and direction of forces influenced the stresses and displacement at the bone–implant interface.
Acknowledgment
This manuscript has been read and approved by all the authors, requirements for authorship have been met, and each author believes that the manuscript represents honest work. We would like to acknowledge Mr. M Nagabhushana for his contribution in the finite element analysis of this study.
Financial support and sponsorship
Nil.
Conflicts of interest
The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or nonfinancial in this article.
References | |  |
1. | Horita S, Sugiura T, Yamamoto K, Murakami K, Imai Y, Kirita T. Biomechanical analysis of immediately loaded implants according to the “All-on-Four” concept. J Prosthodont Res 2017;61:123-32. |
2. | Tribst JP, de Morais DC, Alonso AA, Piva AM, Borges AL. Comparative three-dimensional finite element analysis of implant-supported fixed complete arch mandibular prostheses in two materials. J Indian Prosthodont Soc 2017;17:255-60.  [ PUBMED] [Full text] |
3. | Bhering CL, Mesquita MF, Kemmoku DT, Noritomi PY, Consani RL, Barão VA. Comparison between all-on-four and all-on-six treatment concepts and framework material on stress distribution in atrophic maxilla: A prototyping guided 3D-FEA study. Mater Sci Eng C Mater Biol Appl 2016;69:715-25. |
4. | Najeeb S, Zafar MS, Khurshid Z, Siddiqui F. Applications of polyetheretherketone (PEEK) in oral implantology and prosthodontics. J Prosthodont Res 2016;60:12-9. |
5. | Schwitalla AD, Zimmermann T, Spintig T, Kallage I, Müller WD. Fatigue limits of different PEEK materials for dental implants. J Mech Behav Biomed Mater 2017;69:163-8. |
6. | Esposito M, Hirsch JM, Lekholm U, Thomsen P. Biological factors contributing to failures of osseointegrated oral implants.(II). Etiopathogenesis. Eur J Oral Sci 1998;106:721-64. |
7. | Zoidis P. The all-on-4 modified polyetheretherketone treatment approach: A clinical report. J Prosthet Dent 2018;119:516-21. |
8. | Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-98. |
9. | Li X, Cao Z, Qiu X, Tang Z, Gong L, Wang D. Does matching relation exist between the length and the tilting angle of terminal implants in the all-on-four protocol? Stress distributions by 3D finite element analysis. J Adv Prosthodont 2015;7:240-8. |
10. | Jaros OAL, De Carvalho GAP, Franco ABG, Kreve S, Lopes PAB, Dias SC. Biomechanical behavior of an implant system using polyether ether ketone bar: Finite element analysis. J Int Soc Prev Community Dent 2018;8:446-50. |
11. | Saleh Saber F, Ghasemi S, Koodaryan R, Babaloo A, Abolfazli N. The comparison of stress distribution with different implant numbers and inclination angles in all-on-four and conventional methods in maxilla: A finite element analysis. J Dent Res Dent Clin Dent Prospects 2015;9:246-53. |
12. | Malhotra A, Padmanabhan T, Mohamed K, Natarajan S, Elavia U. Load transfer in tilted implants with varying cantilever lengths in an all-on-four situation. Aust Dent J 2012;57:440-5. |
13. | Lee WT, Koak JY, Lim YJ, Kim SK, Kwon HB, Kim MJ. Stress shielding and fatigue limits of poly-ether-ether-ketone dental implants. J Biomed Biomed Mater Res B 2012;100:1044-52. |
14. | Cicciù M, Bramanti E, Cecchetti F, Scappaticci L, Guglielmino E, Risitano G. FEM and Von Mises analyses of different dental implant shapes for masticatory loading distribution. Oral Implantol (Rome) 2014;7:1-0. |
15. | Winter W, Klein D, Karl M. Micromotion of Dental Implants: Basic Mechanical Considerations. J Med Eng 2013;2013:265412. |
16. | Suedam V, Moretti Neto RT, Sousa EA, Rubo JH. Effect of cantilever length and alloy framework on the stress distribution in peri-implant area of cantilevered implant-supported fixed partial dentures. J Appl Oral Sci 2016;24:114-20. |
17. | van Zyl PP, Grundling NL, Jooste CH, Terblanche E. Three-dimensional finite element model of a human mandible incorporating six osseointegrated implants for stress analysis of mandibular cantilever prostheses. Int J Oral Maxillofac Implants 1995;10:51-7. |
18. | Sugiura T, Yamamoto K, Horita S, Murakami K, Kirita T. Micromotion analysis of different implant configuration, bone density, and crestal cortical bone thickness in immediately loaded mandibular full-arch implant restorations: A nonlinear finite element study. Clin Implant Dent Relat Res 2018;20:43-9. |
19. | Korioth TW, Johann AR. Influence of mandibular superstructure shape on implant stresses during simulated posterior biting. J Prosthet Dent 1999;82:67-72. |
20. | Lee KS, Shin SW, Lee SP, Kim JE, Kim JH, Lee JY. Comparative Evaluation of a Four-Implant-Supported Polyetherketoneketone Framework Prosthesis: A Three-Dimensional Finite Element Analysis Based on Cone Beam Computed Tomography and Computer-Aided Design. Int J Prosthodont 2017;30:581-5. |
21. | Sertgöz A. Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis. Int J Prosthodont 1997;10:19-27. |
[Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
[Table 1], [Table 2], [Table 3], [Table 4], [Table 5]
This article has been cited by | 1 |
Evaluation of stress and strain on mandible caused by changing the bar material in hybrid prosthesis utilizing “All-on-Four” technique |
|
| Yomna H. Shash, Mohamed T. Elwakad, Mohamed A. A. Eldosoky, Mohamed M. Dohiem | | Alexandria Engineering Journal. 2022; | | [Pubmed] | [DOI] | | 2 |
Review on computational modeling for the property, process, product and performance (PPPP) characteristics of additively manufactured porous magnesium implants |
|
| Ramsha Imran, Ans Al Rashid, Muammer Koç | | Bioprinting. 2022; : e00236 | | [Pubmed] | [DOI] | | 3 |
Experimental design of the lower jaw functional loading for implant-supported restoration in unfavorable clinical conditions |
|
| R.A. Rozov, V.N. Trezubov, R.Sh. Gvetadze, A.A. Smerdov, S.O. Chikunov | | Stomatologiya. 2022; 101(6): 28 | | [Pubmed] | [DOI] | | 4 |
Influence of Framework Material and Posterior Implant Angulation in Full-Arch All-on-4 Implant-Supported Prosthesis Stress Concentration |
|
| João Paulo Mendes Tribst, Dayana Campanelli de Morais, Jefferson David Melo de Matos, Guilherme da Rocha Scalzer Lopes, Amanda Maria de Oliveira Dal Piva, Alexandre Luiz Souto Borges, Marco Antonio Bottino, Antonio Lanzotti, Massimo Martorelli, Pietro Ausiello | | Dentistry Journal. 2022; 10(1): 12 | | [Pubmed] | [DOI] | | 5 |
Implant-Supported Prostheses in the Edentulous Mandible: Biomechanical Analysis of Different Implant Configurations via Finite Element Analysis |
|
| Eduardo Anitua, Naiara Larrazabal Saez de Ibarra, Luis Saracho Rotaeche | | Dentistry Journal. 2022; 11(1): 4 | | [Pubmed] | [DOI] | | 6 |
Biomechanical Evaluation of Bone Atrophy and Implant Length in Four Implants Supporting Mandibular Full-Arch-Fixed Dentures |
|
| Heng-Li Huang, Hui-Ling Tsai, Yu-Ling Wu, Jui-Ting Hsu, Aaron Yu-Jen Wu | | Materials. 2022; 15(9): 3295 | | [Pubmed] | [DOI] | | 7 |
Biomechanical investigation of maxillary implant-supported full-arch prostheses produced with different framework materials: a finite elements study |
|
| Mirac Berke Topcu Ersöz, Emre Mumcu | | The Journal of Advanced Prosthodontics. 2022; 14(6): 346 | | [Pubmed] | [DOI] | | 8 |
Evaluation of Stresses on Implant, Bone, and Restorative Materials Caused by Different Opposing Arch Materials in Hybrid Prosthetic Restorations Using the All-on-4 Technique |
|
| Feras Haroun,Oguz Ozan | | Materials. 2021; 14(15): 4308 | | [Pubmed] | [DOI] | |
|
 |
 |
|