Effect of immersion in hydrochloric acid and sodium hypochlorite and autoclave sterilization on the force characteristics of orthodontic nickel-titanium open coils


1 Department of Orthodontics, Dental Research Center, Dental Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran
2 Department of Orthodontics, Student Research Committee, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran

Date of Submission06-Mar-2020
Date of Acceptance10-Sep-2020
Date of Web Publication23-Feb-2021

Correspondence Address:
Dr. Mehrnaz Zakizade
Department of Orthodontics, School of Dentistry, Isfahan University of Medical Science, Isfahan
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1735-3327.310034

Rights and Permissions
  Abstract 


Background: Nickel-titanium (NiTi) open coils are common tools in fixed orthodontic treatments to apply light and continuous forces. Their favorable mechanical properties and their relatively high cost have prompted many clinicians to decide to reuse them.
Aim: The present study aimed to evaluate the effect of 10% hydrochloric acid (HCl) and 5.25% sodium hypochlorite (NaClO) solution and sterilization, on the unloading force of NiTi open-coils.
Materials and Methods: This experimental in vitro study consisted of 160 NiTi open coils from two brands of Highland Metals and Masel were provided. Each group was randomly divided into four subgroups: control, three rounds of immersion, three rounds of autoclave sterilization, and three rounds of immersion and autoclaving. Each round of immersion process included 1 min in 10% HCl, followed by 5 min in 5.25% NaClO. Spring was compressed for 4 mm. Then, the unloading forces of each spring were recorded at 4, 3, 2, and 1 mm of compression, respectively. The data were analyzed with SPSS (P < 0.05), using one-way ANOVA and independent t-test.
Results: One-way ANOVA showed a significant difference in the 2-mm compression of Group 1 and 3-mm compressions of Groups 1 and 2 compared to the control group in the Highland metals brand (P < 0.04, P < 0.014, and P < 0.007, respectively). There was no significant difference between the different compressed lengths and groups in the Masel brand.
Conclusion: One-minute immersion in 10% HCL and 5-min immersion in 5.25% NaOCl, followed by autoclave sterilization, even after repeating the process three times, did not significantly change the force properties of NiTi open springs. Therefore, the re-use of these springs with the above-mentioned method as preparation can be recommended.

Keywords: Hydrochloric acid, orthodontics, recycling, sodium hypochlorite


How to cite this article:
Noorollahian S, Zakizade M. Effect of immersion in hydrochloric acid and sodium hypochlorite and autoclave sterilization on the force characteristics of orthodontic nickel-titanium open coils. Dent Res J 2021;18:5

How to cite this URL:
Noorollahian S, Zakizade M. Effect of immersion in hydrochloric acid and sodium hypochlorite and autoclave sterilization on the force characteristics of orthodontic nickel-titanium open coils. Dent Res J [serial online] 2021 [cited 2021 May 11];18:5. Available from: https://www.drjjournal.net/text.asp?2021/18/1/5/310034




  Introduction Top


Nickel-titanium (NiTi) alloy has been used in orthodontics since the early 1970s.[1] Due to the proper mechanical properties of this alloy, its use has gradually increased, and it has won great popularity in recent decades.[2] The positive characteristic of this alloy is related to its low load deflection rate and, consequently, high flexibility and elasticity.[3]

NiTi springs are a common tool in fixed orthodontic treatments to apply light and continuous forces. These springs have been manufactured in both open and closed forms.[4],[5] Open coils are used to provide the space for impacted teeth, severely mal-aligned teeth, distalization of molars, and retraction of incisors.[6],[7] NiTi springs apply the more appropriate force in a longer activation range compared to steel springs. Thus, the force applied is more physiologic.[5],[8]

NiTi alloys are expensive.[9] Their favorable mechanical properties and their relatively high cost have prompted many clinicians to decide to re-use them.[10] Economic considerations are the most important cause of their reuse.[11],[12] Their manufacturing process needs to use limited natural resources; therefore, their reuse could help correct this imbalance in the environment.[13],[14] By reusing, costs are reduced by 30%–40% and nature is protected against the depletion of resources.[13]

Reuse of orthodontic appliances is possible when, in addition to preventing the transmission of infections, the physical characteristics of the appliance do not significantly change after the initial application, cleaning, disinfection, and sterilization processes.[15]

It is possible to reuse tools and appliances that retain their function after clinical use by observing some principles. This is also implemented in medicine and other dental fields. The reuse of some appliances has also been suggested in orthodontics. These include brackets,[16],[17],[18],[19] bands,[20] wires,[21],[22],[23],[24] orthodontic mini-implants.[25],[26]

A literature search did not bring up a study on the reuse of NiTi open coils. However, some studies have evaluated the reuse of closed coils. Momeni et al. showed that autoclaving could increase their strength; however, the increase was not clinically significant. They claimed that Niti close coils could be reused without a significant change in their strength.[27] Geng et al. reported that thermocycling resulting from the daily use of foods and drinks might result in the loss of the force of NiTi close coils.[28] Magno et al. showed that 6-month clinical use of NiTi close coils could result in a significant increase in stress-strain curve, a decrease in strength, and permanent deformation.[29]

The results on close coils are contradictory and needed more researches. It should be noted that the manufacturing conditions and distribution of force due to deformation in open and closed coils are different. There are limited studies available on the reuse of NiTi open coils. Due to their favorable characteristics and high cost, reusing of these springs can be suggested. The present study is aimed to evaluate the effect of the use of 10% hydrochloric acid (HCL) and 5.25% sodium hypochlorite (NaClO) solution to eliminate mineral and organic deposits on NiTi open coils, respectively, followed by sterilization, on unloading force of these springs.


  Materials and Methods Top


Ethical approval was granted by the Torabinejad Dental Research Center, Isfahan University of Medical Sciences, Isfahan, Iran, with identification number 398835. This experimental in vitro study consists of 160 orthodontic NiTi open coils, 14 mm in length, including 80 from Highland Metals (Encinitas, California, USA) and 80 from Masel) San Jose, California, USA) brand. Each group was randomly divided into four subgroups. By setting a confidence level at 0.95, power at 0.80, and the standard deviation at 6, the sample size should be 20 samples for each group. Groups 0 were control, and Groups 1–3 underwent three rounds of immersion, three rounds of autoclave sterilization, and three rounds of immersion and autoclaving, respectively. Each round of immersion process included 1 min in 10% HCl (Household cleaner solution, Golrang Industrial Group, Tehran, Iran), followed by 10 s irrigation with running water and followed by 5 min in 5.25% NaClO (Bleaching solution, Golrang Industrial Group, Tehran, Iran). Each sample was immersed in a 20 mL fresh solution separately. The autoclaving process included 10 min in 121°C temperature and +1 bar pressure (Gemini BV, Apeldoorn, Netherland).

After this step, the first researcher coded the samples and the second researcher measured the spring unloading forces using an electromechanical universal testing machine (K-21046, Water + Bai, Switzerland) at room temperature [Figure 1]. Spring was compressed up to 4 mm. Then, the unloading forces of each spring were recorded at 4, 3, 2, and 1 mm of compression, respectively. The data after decoding were analyzed with SPSS (version 22) SPSS (version 22) (SPSSInc., Chicago, IL,USA) (P < 0.05), using one-way ANOVA and independent t-test. The level of significance was set at 0.05. The normality was analyzed using Kolmogorov–Smirnov test.
Figure 1: Measuring unloading force of each coil.

Click here to view



  Results Top


Kolmogorov–Smirnov test showed the normal distribution of data. One-way ANOVA analysis was carried out to compare the Groups at 1, 2, 3, and 4 mm of activation separately. The mean levels of force at different compressions are shown in [Table 1].
Table 1: Mean (standard deviation) of unloading forces (cN) of nickel-titanium open coil springs in different groups and compression lengths

Click here to view


In the Highland metals brand, there were no significant differences between groups except in the 2-mm compression of Group 1 and 3-mm compressions of Groups 1 and 2 compared to the control group. (P < 0.04, P < 0.014, and P < 0.007, respectively) There was no significant difference between the different groups in the Masel brand. The control groups of the two brands compared with independent t-test. Masel springs had higher unloading forces than Highlands springs at 1 and 2 mm compression (P < 0.001, P = 0.017) but at 3 and 4 mm compression there were no significant differences (P < 0.066, P = 0.972).


  Discussion Top


The most important rationale of medical and dental instruments reuse is economic issues.[11],[12] The removal of organic and mineral debris as cleaning is necessary before sterilization. In this study, immersion in 10% HCL, followed by immersion in 5.25% NaOCl, was proposed as a method for cleaning NiTi open coils before autoclave sterilization. They are known as household cleaner and bleaching solution, and they are inexpensive and available. HCL can remove mineral debris.[16] NaOCl can resolve organic debris and is a common and powerful disinfectant[25],[26] Autoclave sterilization is carried out as a routine process of sterilization. Cleaning and sterilization processes must not have an adverse effect on the physical properties of equipment if they are reused.[30],[31]

In this study, two different brands of NiTi open coils were used to help better the generalization of the results. Measuring unloading forces at 1–4-mm compression is done to simulate routine clinical use of these coils.

Three times immersion did not show significant changes in force characteristics of the Masel coils; however, the immersion procedure with Highland Metals coils at 2 and 3 mm compression resulted in significant changes, statistically, in force characteristics up to 12.6 and 20 cN, respectively, which is not significant clinically.

Three rounds of autoclave sterilization in the Masel group did not result in significant changes in force levels. However, concerning the Highland Metals coils, a decrease of 21.54 cN was recorded in force only at 3 mm compression, with no significant differences at other points. This amount of decrease in force is not significant clinically. Momeni et al.[27] found that even though autoclave sterilization can increase the force levels of Ni-Ti close coils but this is not clinically significant, and it is possible to recycle NiTi closed-coil springs without a considerable reduction in their force levels.

No studies were found about the effect of reuse on NiTi open coils. However, several studies have investigated the effects of clinical use and sterilization on NiTi close coils and NiTi wires. Several studies have supported the reuse of NiTi close coils.[2],[27],[32],[33]

Crotty examined the effect of autoclave on tensile properties of NiTi wires and reported no significant changes in the tensile properties of NiTi wires sterilized in an autoclave,[34] which were consistent with Lee's findings.[35] On the other hand, Bavikati et al. showed that autoclave sterilization of NiTi wires and clinical reuse of the wires could reduce the elasticity of the NiTi wires and affect their surface topography.[36] Alavi also reported the changes in the mechanical properties of NiTi wires after autoclaving.[9] There was no significant decrease in the force of any of the brands by concomitant autoclaving and immersion for three rounds. No similar study was found for comparison concerning the effect of the cleaning/sterilization process on the spring force.

Three times repeating the process of immersion and sterilization in this study were used to investigate the possibility of reuse for several times. By accepting the lack of the effect of the above process three times on the spring properties, the results could be generalized with greater confidence to less repetitions of the reuse.


  Conclusion Top


One-minute immersion in 10% HCL and 5-min immersion in 5.25% NaOCl, followed by autoclave sterilization, even after repeating the process three times, did not significantly change the force properties of NiTi open springs. Therefore, the reuse of these coils with the above-mentioned method can be recommended.

Financial support and sponsorship

Nil.

Conflicts of interest

The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or non-financial in this article.



 
  References Top

1.
Han S, Quick DC. Nickel-titanium spring properties in a simulated oral environment. Angle Orthod 1993;63:67-72.  Back to cited text no. 1
    
2.
Wichelhaus A, Brauchli L, Ball J, Mertmann M. Mechanical behavior and clinical application of nickel-titanium closed-coil springs under different stress levels and mechanical loading cycles. Am J Orthod Dentofacial Orthop 2010;137:671-8.  Back to cited text no. 2
    
3.
Kapila S, Reichhold GW, Anderson RS, Watanabe LG. Effects of clinical recycling on mechanical properties of nickel-titanium alloy wires. Am J Orthod Dentofacial Orthop 1991;100:428-35.  Back to cited text no. 3
    
4.
Samuels RH, Peak JD. Use of nickel titanium closed-coil springs to align unerupted teeth: A case report. Am J Orthod Dentofacial Orthop 1998;113:353-8.  Back to cited text no. 4
    
5.
von Fraunhofer JA, Bonds PW, Johnson BE. Force generation by orthodontic coil springs. Angle Orthod 1993;63:145-8.  Back to cited text no. 5
    
6.
Brauchli LM, Senn C, Ball J, Wichelhaus A. Force levels of 23 nickel-titanium open-coil springs in compression testing. Am J Orthod Dentofacial Orthop 2011;139:601-5.  Back to cited text no. 6
    
7.
Petrov A, Davidyan A, Goonewardene MS. Incisor retraction with a compressed open-coil spring. J Clin Orthod 2014;48:653.  Back to cited text no. 7
    
8.
Santoro M, Nicolay OF, Cangialosi TJ. Pseudoelasticity and thermoelasticity of nickel-titanium alloys: A clinically oriented review. Part II: Deactivation forces. Am J Orthod Dentofacial Orthop 2001;119:594-603.  Back to cited text no. 8
    
9.
Alavi SR, Ghorbani AA. Effects of steam and dry-heat sterilization on bending properties of NiTi wires. Orthod Waves 2009;68:123-8.  Back to cited text no. 9
    
10.
Mayhew MJ, Kusy RP. Effects of sterilization on the mechanical properties and the surface topography of nickel-titanium arch wires. Am J Orthod Dentofacial Orthop 1988;93:232-6.  Back to cited text no. 10
    
11.
Manatakis DK, Georgopoulos N. Reducing the cost of laparoscopy: Reusable versus disposable laparoscopic instruments. Minim Invasive Surg 2014;2014:1-4.  Back to cited text no. 11
    
12.
Unger SL. Assessing the environmental, human health, and economic impacts of reprocessed medical devices in a Phoenix hospital's supply chain. J Clean Prod 2016;112:1995-2003.  Back to cited text no. 12
    
13.
Thopegowda N, Shenoy K, Shankarnarayana R, Kukkila J, Vaddya S, Ginjupalli K. Recycling of Materials used in Dentistry with Reference to its Economical and Environmental Aspects. Int J Health Rehabil Sci 2013;2:140-5.  Back to cited text no. 13
    
14.
Grantcharov P, Ahmed S, Wac K, Rivas H. Reprocessing and reuse of single-use medical devices: Perceptions and concerns of relevant stakeholders toward current practices. Int J Evid Based Healthc 2019;17:53-7.  Back to cited text no. 14
    
15.
Montgomery SR. Disposable versus reusable blood pressure cuffs: A nursing led initiative. J Perioper Crit Intensive Care Nurs 2016;2:1-3.  Back to cited text no. 15
    
16.
Bahnasi FI, Abd-Rahman AN, Abu-Hassan MI. Effects of recycling and bonding agent application on bond strength of stainless steel orthodontic brackets. J Clin Exp Dent 2013;5:e197-202.  Back to cited text no. 16
    
17.
Han RQ, Yang K, Ji LF, Ling C. Analysis of shear bond strength and morphology of Er: YAG laser-recycled ceramic orthodontic brackets. Biomed Res Int 2016;2016:1-6.  Back to cited text no. 17
    
18.
Yassaei S, Aghili H, Hosseinzadeh Firouzabadi A, Meshkani H. Effect of Er: YAG laser and sandblasting in recycling of ceramic brackets. J Lasers Med Sci 2017;8:17-21.  Back to cited text no. 18
    
19.
Kachoei M, Mohammadi A, Esmaili Moghaddam M, Rikhtegaran S, Pourghaznein M, Shirazi S. Comparison of multiple rebond shear strengths of debonded brackets after preparation with sandblasting and CO2 laser. J Dent Res Dent Clin Dent Prospects 2016;10:148-54.  Back to cited text no. 19
    
20.
Fulford MR, Ireland AJ, Main BG. Decontamination of tried-in orthodontic molar bands. Eur J Orthod 2003;25:621-2.  Back to cited text no. 20
    
21.
Oshagh M, Hematiyan MR, Mohandes Y, Oshagh MR, Pishbin L. Autoclaving and clinical recycling: Effects on mechanical properties of orthodontic wires. Indian J Dent Res 2012;23:638-42.  Back to cited text no. 21
[PUBMED]  [Full text]  
22.
Pernier C, Grosgogeat B, Ponsonnet L, Benay G, Lissac M. Influence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires. Eur J Orthod 2005;27:72-81.  Back to cited text no. 22
    
23.
Khier SE, Brantley WA, Fournelle RA. Structure and mechanical properties of as-received and heat-treated stainless steel orthodontic wires. Am J Orthod Dentofacial Orthop 1988;93:206-12.  Back to cited text no. 23
    
24.
Ramazanzadeh BA, Ahrari F, Sabzevari B, Zebarjad SM, Ahrari A. Effects of a simulated oral environment and sterilization on load-deflection properties of superelastic nickel titanium-based orthodontic wires. Int J Orthod Milwaukee 2011;22:13-21.  Back to cited text no. 24
    
25.
Noorollahian S, Alavi S, Monirifard M. A processing method for orthodontic mini-screws reuse. Dent Res J (Isfahan) 2012;9:447-51.  Back to cited text no. 25
    
26.
Estelita S, Janson G, Chiqueto K, Ferreira ES. Effect of recycling protocol on mechanical strength of used mini-implants. Int J Dent 2014;2014:1-5.  Back to cited text no. 26
    
27.
Momeni Danaei S, Oshagh M, Khozaei A. Effect of recycling and autoclave sterilization on the unloading forces of NiTi closed-coil springs: An in vitro study. J Dent (Shiraz) 2013;14:184-90.  Back to cited text no. 27
    
28.
Geng H, Su H, Whitley J, Lin FC, Xu X, Ko CC. The effect of orthodontic clinical use on the mechanical characteristics of nickel-titanium closed-coil springs. J Int Med Res 2019;47:803-14.  Back to cited text no. 28
    
29.
Magno AF, Monini Ada C, Capela MV, Martins LP, Martins RP. Effect of clinical use of nickel-titanium springs. Am J Orthod Dentofacial Orthop 2015;148:76-82.  Back to cited text no. 29
    
30.
Eckelman M, Mosher M, Gonzalez A, Sherman J. Comparative life cycle assessment of disposable and reusable laryngeal mask airways. Anesth Analg 2012;114:1067-72.  Back to cited text no. 30
    
31.
Sloan T. First, do no harm a framework for evaluating new versus reprocessed medical devices. J Ope Res Soc 2010;61:191-201.  Back to cited text no. 31
    
32.
Vidoni G, Perinetti G, Antoniolli F, Castaldo A, Contardo L. Combined aging effects of strain and thermocycling on unload deflection modes of nickel-titanium closed-coil springs: An in vitro comparative study. Am J Orthod Dentofacial Orthop 2010;138:451-7.  Back to cited text no. 32
    
33.
Alavi S, Haerian A. The effects of aging process and preactivation on mechanical properties of nickel-titanium closed coil springs. Dent Res J (Isfahan) 2015;12:231-4.  Back to cited text no. 33
    
34.
Crotty OP, Davies EH, Jones SP. The effects of cross-infection control procedures on the tensile and flexural properties of superelastic nickel-titanium wires. Br J Orthod 1996;23:37-41.  Back to cited text no. 34
    
35.
Lee SH, Chang YI. Effects of recycling on the mechanical properties and the surface topography of nickel-titanium alloy wires. Am J Orthod Dentofacial Orthop 2001;120:654-63.  Back to cited text no. 35
    
36.
Bavikati VN, Singaraju GS, Mandava P, Killamsetty SS, Nettam V, Karnati PK. Evaluation of mechanical and physical properties of clinically used and recycled superelastic NiTi Wires. J Clin Diagn Res 2016;10:ZC35-40.  Back to cited text no. 36
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1]



 

Top
 
 
Search
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed421    
    Printed4    
    Emailed0    
    PDF Downloaded53    
    Comments [Add]    

Recommend this journal