Dental Research Journal

ORIGINAL ARTICLE
Year
: 2018  |  Volume : 15  |  Issue : 5  |  Page : 347--353

The effect of microthread design on magnitude and distribution of stresses in bone: A three-dimensional finite element analysis


Shima Golmohammadi1, Amir Eskandari2, Mohammad Reza Movahhedy3, Adileh Shirmohammadi2, Reza Amid4 
1 Department of Periodontics, Faculty of Dentistry, Islamic Azad University of Borujerd, Borujerd, Iran
2 Department of Periodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran
3 Department of Biomechanics, Faculty of Mechanical Engineering, Sharif University of Technology, Tehran, Iran
4 Department of Periodontics, Faculty of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Correspondence Address:
Dr. Shima Golmohammadi
Department of Periodontics, Faculty of Dentistry, Lorestan University of Medical Sciences, Khorramabad, Lorestan
Iran

Background: The researches regarding the influence of microthread design variables on the stress distribution in bone and a biomechanically optimal design for implant neck are limited. The aim of the present study is to compare the effect of different microthread designs on crestal bone stress. Materials and Methods: Six implant models were constructed for three-dimensional finite element analysis including two thread profile (coarse and fine) with three different lengths of microthreaded neck (1 mm, 2 mm, and 3 mm). A load of 200 N was applied in two angulations (0° and 30°) relative to the long axis of the implant and the resultant maximum von Mises equivalent (EQV), compressive, tensile, and shear stresses were measured. Results: Regardless of loading angle, the highest EQV stress was concentrated in the cortical bone around the implant model using a 1 mm neck of fine microthreads. Under axial loading, there was a negative correlation between the length of the microthreaded neck and stress level in both profiles. However, the same pattern was not observed for coarse microthreads under oblique loads. All types of measured stresses in all constructed models were increased with oblique loading. Conclusion: Peak stress levels in implant models varied with microthread profile and direction of loading. The microthread profile seemed more important than the length of the neck in reducing loading stresses exerted on the surrounding bone. Fine microthreads on a 3 mm implant neck showed consistently higher cortical bone stress than other models.


How to cite this article:
Golmohammadi S, Eskandari A, Movahhedy MR, Shirmohammadi A, Amid R. The effect of microthread design on magnitude and distribution of stresses in bone: A three-dimensional finite element analysis.Dent Res J 2018;15:347-353


How to cite this URL:
Golmohammadi S, Eskandari A, Movahhedy MR, Shirmohammadi A, Amid R. The effect of microthread design on magnitude and distribution of stresses in bone: A three-dimensional finite element analysis. Dent Res J [serial online] 2018 [cited 2020 Oct 20 ];15:347-353
Available from: https://www.drjjournal.net/article.asp?issn=1735-3327;year=2018;volume=15;issue=5;spage=347;epage=353;aulast=Golmohammadi;type=0